Application of a cw quantum cascade laser CO2 analyser to catalytic oxidation reaction monitoring
Authors:V L Kasyutich  D Poulidi  M Jalil  I S Metcalfe  P A Martin
Institution:1. School of Chemical Engineering and Analytical Science, University of Manchester, Oxford Road, Manchester, M13 9PL, UK
2. School of Chemical Engineering and Advanced Materials, Metz Court, Newcastle University, Newcastle upon Tyne, NE1 7RU, UK
Abstract:Catalytic oxidation reaction monitoring has been performed for the first time with a trace gas carbon dioxide analyser based on a continuous wave (cw), thermoelectrically cooled (TEC), distributed feedback (DFB) quantum cascade laser (QCL) operating at around 2307 cm?1. The reaction kinetics for carbon monoxide oxidation over a platinum catalyst supported on yttria-stabilised zirconia were followed by the QCL CO2 analyser and showed that it is a powerful new tool for measuring low reaction rates associated with low surface area model catalysts operating at atmospheric pressures. A detection limit was determined of 40 ppb (1 standard deviation) for a 0.1 s average and a residual absorption standard deviation of 1.9×10?4.