Detector area expansion at iNSE neutron spin echo spectrometer
Authors:NL Yamada  H Endo  N Osaka  Y Kawabata  M Nagao  T Takeda  H Seto  M Shibayama
Institution:aNeutron Science Laboratory, High Energy Accelerator Research Organization, Tsukuba 305-0801, Japan;bInstitute for Solid State Physics, The University of Tokyo, Tokai 319-1106, Japan;cDepartment of Chemistry, Faculty of Science, Tokyo Metropolitan University, Tokyo 192-0397, Japan;dGraduate School of Integrated Arts and Sciences, Hiroshima University, Higashihiroshima 739-8521, Japan
Abstract:Expansion of a detection area is an effective method to increase the measurement efficiency of a neutron spin echo (NSE) spectrometer as well as other spectrometers. For this purpose, we installed a new π/2 spin flipper and Fresnel coil of iNSE spectrometer at JRR-3, Tokai, Japan, for wide-area data acquisition. In this study, we propose a data reduction method to correct the phase inhomogeneity due to the path difference of neutrons on the large detection area. This method can convert many NSE signals at small areas to one NSE signal at a large area with taking the phase offset due to the phase inhomogeneity into account. The measurement efficiency increased by approximately one order of magnitude as a result of the detection area expansion.
Keywords:Neutron spin echo (NSE)  Detector area expansion  Phase correction method
本文献已被 ScienceDirect 等数据库收录!