首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A comparative study of systems used for dynamic focusing of ultrasound
Authors:A P Sarvazyan  L Fillinger  L R Gavrilov
Institution:1. Artann Laboratories, Trenton, NJ, 08618, USA
2. Andreev Acoustics Institute, Russian Academy of Sciences, ul. Shvernika 4, Moscow, 117036, Russia
Abstract:A comparative study of two methods used for dynamic focusing of ultrasound: the conventional phased arrays and a new method based on time reversal of acoustic signals is carried out. A laboratory model of the focusing system based on time reversal is developed and manufactured. One of the principal elements of the system is a reverberator with several piezoelectric transducers attached to its walls. Experiments are carried out to demonstrate the ability of such a system to generate one focus or several foci and to steer them electronically at considerable distances (50 mm at minimum) off the axis of the focusing system without causing the appearance of any grating lobes or other secondary intensity maxima. The focusing properties of the system are compared with the results of numerical simulation of two-dimensional phased arrays, whose parameters are taken to be typical for the arrays used in extracorporeal surgery. The important role of randomization is demonstrated for both of the aforementioned focusing methods. The prospects of practical application of the two methods are discussed.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号