Evaluation of aortic stenosis severity using 4D flow jet shear layer detection for the measurement of valve effective orifice area
Authors:Julio Garcia  Michael Markl  Susanne Schnell  Bradley Allen  Pegah Entezari  Riti Mahadevia  S Chris Malaisrie  Philippe Pibarot  James Carr  Alex J Barker
Institution:1. Radiology, Northwestern University, Chicago, USA;2. Medicine, Laval University, Quebec, Canada;3. Biomedical Engineering, Northwestern University, Chicago, USA;4. Division of Cardiothoracic Surgery, Northwestern University, Chicago, USA
Abstract:

Aims

The objective of this study was to evaluate the potential of 4D flow MRI to assess valve effective orifice area (EOA) in patients with aortic stenosis as determined by the jet shear layer detection (JSLD) method.

Methods and Results

An in-vitro stenosis phantom was used for validation and in-vivo imaging was performed in 10 healthy controls and 40 patients with aortic stenosis. EOA was calculated by the JSLD method using standard 2D phase contrast MRI (PC-MRI) and 4D flow MRI measurements (EOAJSLD-2D and EOAJSLD-4D, respectively). As a reference standard, the continuity equation was used to calculate EOA (EOACE) with the 2D PC-MRI velocity field and compared to the EOAJSLD measurements. The in-vitro results exhibited excellent agreement between flow theory (EOA = 0.78 cm2) and experimental measurement (EOAJSLD-4D = 0.78 ± 0.01 cm2) for peak velocities ranging from 0.9 to 3.7 m/s. In-vivo results showed good correlation and agreement between EOAJSLD-2D and EOACE (r = 0.91, p < 0.001; bias: − 0.01 ± 0.38 cm2; agreement limits: 0.75 to − 0.77 cm2), and between EOAJSLD-4D and EOACE (r = 0.95, p < 0.001; bias: − 0.09 ± 0.26 cm2; limits: 0.43 to − 0.62 cm2).

Conclusion

This study demonstrates the feasibility of measuring EOAJSLD using 4D flow MRI. The technique allows for optimization of the EOA measurement position by visualizing the 3D vena contracta, and avoids potential sources of EOACE measurement variability.
Keywords:4D flow MRI  Aortic stenosis  Effective orifice area  Jet shear layer