Using a plane wave approximation in simulation of radiation distribution in an X-ray Talbot interferometer
Authors:A S Gogolev  M A Kazaryan  A V Obkhodsky  A S Popov  R O Rezaev  E A Smirnova
Institution:1.National Research Tomsk Polytechnic University,Tomsk,Russia;2.Lebedev Physical Institute,Russian Academy of Sciences,Moscow,Russia;3.National Research Nuclear University “MEPhI”,Moscow,Russia
Abstract:This study is devoted to the analysis of the plane-wave approximation applicability to X-ray radiation incident on an object. Based on simple calculations, it is shown that an X-ray tube focal spot 0.4 × 0.8 mm in size at distances of ~1 m can be considered as a point source; however, the plane-wave approximation for such X-ray radiation with energy E = 22keV is valid for radiation source–object distances much longer than 10 m.