首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Measurement of indoor Rn using CR-39 under a thin film geometry
Authors:R S Neman  I D Schmitman  J C Hadler N  P J Iunes  S R Paulo  S Guedes
Institution:

a Instituto de Física “Gleb Wataghin”, Universidade Estadual de Campinas, UNICAMP, 13083-970 Campinas, SP, Brazil

b Departamento de Física, Instituto de Ciências Exatas e da Terra, Universidade Federal de Mato Grosso, UFMT, 78060-900, Cuiaba, MT, Brazil

Abstract:In this work an assembly for indoor 222Rn measurement is presented. This assembly is made up of two acrylic plates (14 cm×14 cm) separated by a distance of 4 mm. To prevent radon daughters from outside entering the assembly (and produce alpha particle tracks striking the CR-39 detector placed in the central region of one of these plates), the borders of these plates were progressively closed, leaving open only a rectangular aperture of 5 mm×4 mm. The size of this opening was determined experimentally, by exposing in two indoor environments assemblies with different apertures as follows: (i) all borders open; (ii) two borders closed; (iii) three borders closed; (iv) four borders closed but one containing a 6 cm×4 mm opening; (v) four borders closed but one containing a 2.5 cm×4 mm opening and (vi) four borders closed but one containing a 0.5 cm×4 mm opening. Track density shows a noticeable decrease between assembly (i) and assembly (iii), remaining constant for smaller openings. Only 222Rn, a noble gas, should enter the assembly independently of the opening size.
Keywords:Radon  Measurement assembly  CR-39
本文献已被 ScienceDirect 等数据库收录!