首页 | 本学科首页   官方微博 | 高级检索  
     检索      

应用蒙特卡罗模拟进行正电子发射断层成像仪散射特性分析
引用本文:高飞,山田亮子,渡边光男,刘华锋.应用蒙特卡罗模拟进行正电子发射断层成像仪散射特性分析[J].物理学报,2009,58(5):3584-3591.
作者姓名:高飞  山田亮子  渡边光男  刘华锋
作者单位:(1)日本滨松光子学株式会社中央研究所,滨松,日本 434-8601; (2)浙江大学现代光学仪器国家重点实验室,杭州 310027
基金项目:国家自然科学基金(批准号:60772125)、广东省重点基金(批准号:07117784)和浙江省科技计划(批准号:2008C23060)资助的课题
摘    要:通过使用基于Geant4的蒙特卡罗模拟代码GATE对全身用全3D正电子发射断层成像仪和含有挡板的2D/3D小动物用正电子发射断层成像仪进行建模,系统地分析了3D采集条件下正电子发射断层成像仪的散射分数、散射分布、多次散射、视野外散射四个主要方面和2D采集条件下挡板对散射分数和散射分布的影响.针对全3D散射校正的难点: 多次散射和视野外散射,设计了附加实验,拟合得到了多次散射光子的百分比随体模横截面积变化的关系和不同环的位置受到视野外散射光子的影响;针对2D散射校正,对挡板引入的散射光子进行分离,单独分析, 关键词: 正电子发射断层成像仪(PET) 蒙特卡罗模拟 散射特性 散射校正

关 键 词:正电子发射断层成像仪(PET)  蒙特卡罗模拟  散射特性  散射校正
收稿时间:2008-08-26

Use of Monte Carlo simulations for the scatter events analysis of PET scanners
Gao Fei,Ryoko Yamada,Mitsuo Watanabe,Liu Hua-Feng.Use of Monte Carlo simulations for the scatter events analysis of PET scanners[J].Acta Physica Sinica,2009,58(5):3584-3591.
Authors:Gao Fei  Ryoko Yamada  Mitsuo Watanabe  Liu Hua-Feng
Abstract:Monte Carlo simulation is an essential tool to assist in the design of new medical imaging devices for emission tomography. This paper modeled the full 3D whole body PET scanner and 2D animal PET scanner with septa using Monte Carlo simulation code GATE based on Geant4, analyzed the scatter fraction, scatter distribution, multiple scatter events, and scatter events from outside of FOV under 3D acquisition mode and scatter events from the septa under 2D acquisition mode. For the critical problems of 3D scatter correction: multiple scatter events and scatter events from outside of FOV, we designed additional experiments, got the fitted relationship between the percentage of multiple scatter events and the cross sectional area of the phantom, and the effects of the scatter events from outside of FOV at different ring positions; for the 2D scatter correction, we separated the scatter events from septa for analysis, and provided a quantitative reference for the elimination of these events or introduction into the procedure of reconstruction.
Keywords:positron emission tomography (PET)  Monte Carlo simulation  scatter characters  scatter correction
本文献已被 万方数据 等数据库收录!
点击此处可从《物理学报》浏览原始摘要信息
点击此处可从《物理学报》下载免费的PDF全文