超巨磁阻测辐射热仪
引用本文:顾梅梅,张鹏翔,李国桢.超巨磁阻测辐射热仪[J].物理学报,2000,49(8):1567-1573.
作者姓名:顾梅梅  张鹏翔  李国桢
作者单位:(1)昆明理工大学材料系,昆明 650051; (2)昆明理工大学材料系,昆明 650051;昆明物理研究所超导组,昆明 650223
基金项目:云南省科学技术委员会重点项目(批准号:YKS199907022)、云南省科学技术委员会项目(批准号:98E056M)资助的课题.
摘    要:基于超巨磁阻(CMR)材料在金属-绝缘体(M-I)转变点附近的巨大电阻变化,设计并测试了一种新型超巨磁阻测辐射热仪(CMR bolometer).用外延法生长的La0.67Ca0.33MnO3 薄膜作为测辐射热仪的辐射敏感元件,测量了该器件对黑体和He-Ne激光器的光学响应.所测的信号、噪声都随调制频率的上升而下降,但对于He-Ne激光源,它的信噪比却没有明显的改变.测量了薄膜铁磁金属态和顺磁绝缘态的信号-温度关系,最强的信号出现在靠近M- 关键词: 超巨磁阻

关 键 词:超巨磁阻
收稿时间:1/4/2000 12:00:00 AM

COLOSSAL MAGNETORESISTANCE BOLOMETER
GU MEI-MEI,ZHANG PENG-XIANG,LI GUO-ZHEN.COLOSSAL MAGNETORESISTANCE BOLOMETER[J].Acta Physica Sinica,2000,49(8):1567-1573.
Authors:GU MEI-MEI  ZHANG PENG-XIANG  LI GUO-ZHEN
Abstract:Based on the sharp resistance change near the metal-insulator transition (M-I) of the colossal magnetoresistance (CMR) materials, we design and test a new type CMR bolometer. La0.67Ca0.33MnO3 epitaxial thin films were used as the photosensitive element of the bolometer, and the optical response of this device to the He-Ne laser and the blackbody were measured. The measured signal and noise decrease with increasing modulation frequency; however, the signal-to-noise ratio shows no obvious change for the He-Ne laser. The temperature dependence of signal was measured in both ferromagnetic metallic and paramagnetic insulating states. It shows that the strongest signal appears in the ferromagnetic state very close to the M-I point. The film's Curie temperature can span from 70 to 350K, depending on the manufacture conditions. So, working near or at room temperature characterizes this CMR thin film bolometer. In this sense, it compensates the weakness of the high-temperature superconductor bolometer that can be used only around the liquid Nitrogen (LN) temperature.The measured detectivity is about 107 cmHz1/2/W and the thermal response constant is 20ms for this device. Compared with the high-temperature superconductor bolometer, the time constant of CMR bolometer is longer and the detectivity is lower. However, there is much room to improve yet.
Keywords:colossal magnetoresistance  bolometer