首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Characterization of a triple-GEM position sensitive detector for X-ray fluorescence imaging
Authors:Geovane G A de Souza  Hugo Natal da Luz
Institution:Department of Nuclear Physics, Instituto de Física da Universidade de São Paulo, São Paulo, Brasil
Abstract:In this work, we characterized an X-ray position sensitive gaseous detector based in a triple stack of gas electron multipliers. The readout circuit is divided in 256 strips for each dimension and using a resistive chain interconnecting the strips, we are able to reconstruct the radiation interaction points by resistive charge division. The detector achieved gains above 104, energy resolution of 15.28% (full width at half maximum) for 5.9 keV X-rays, and position resolution of 1.2 mm, while operating in Ar/CO2(90/10) at atmospheric pressure.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号