Filter-radiometer-based realization of candela and establishment of photometric scale at UME |
| |
Authors: | Farhad Samedov Murat Durak
zcan Bazkr |
| |
Institution: | TÜBİTAK-Ulusal Metroloji Enstitüsü (UME) 41470, Gebze, Kocaeli, Turkey |
| |
Abstract: | The luminous intensity unit of candela was realized based on filter-radiometer, which is traceable to detector-based primary standard electrical substitution cryogenic radiometer (ESCR). In that realization the traditional Osram Wi41/G-type incandescent lamp and filter-radiometer consisting of an aperture, a V(λ) filter and a silicon photodiode based trap detector were used as light source and detection element, respectively. Measurement techniques of effective aperture area, spectral transmittance of V(λ) filter and absolute responsivity of trap detector are also presented. |
| |
Keywords: | Responsivity Optical power Filter-radiometer Trap detector Luminous intensity Candela |
本文献已被 ScienceDirect 等数据库收录! |