High precision automated conductometric titrations with the bipolar pulse technique |
| |
Authors: | Mark E Hail F James Holler |
| |
Institution: | (1) Department of Chemistry, University of Kentucky, 40506 Lexington, KY, USA;(2) Present address: Department of Chemistry, University of Florida, 32611 Gainesville, FL, USA |
| |
Abstract: | An automated conductometric titration system has been designed and investigated. Stepper motor driven burets have been constructed to provide precise delivery of small volumes of titrant. A bipolar pulse conductance instrument of wide dynamic range is briefly reviewed as a tool in conductometric titrations. Several types of titrations have been carried out to investigate the accuracy and reproducibility of this apparatus. The precision for titrant volumes of 40µl is ±0.64% relative standard deviation. |
| |
Keywords: | titration conductance automation computer control |