Azo calix[4]arene based neodymium(III)-selective PVC membrane sensor |
| |
Authors: | Menon Shobhana K Modi Nishith R Patel Bhargav Patel Manishkumar B |
| |
Institution: | a Department of Chemistry, School of Sciences, Gujarat University, Navrangpura, Ahmedabad 380009, Gujarat, India |
| |
Abstract: | We found that the PVC membrane, containing azo calix4]arene is a suitable ionophore, exhibited a Nernstian response for neodymium (Nd3+) ions (with slope of 19.8 ± 0.2 mV decade?1 for the triply charged ion) over a wide linear range of 4.0 × 10?8 to 1.0 × 10?1 mol L?1 with a detection limit 1.0 × 10?8 mol L?1, a relatively fast response time, in the whole concentration range (<10 s), and a considerable life time at least for four months in the pH range of 4.0-8.0. Furthermore, the electrode revealed high selectivity with respect to all the common alkali, alkaline earth, transition and heavy metal ions, including the members of the lanthanide family other than Nd3+. Concerning its applications, it was effectively employed for the determination of neodymium ions in industrial waste water as well as in lake water. |
| |
Keywords: | Azo calix[4]arene Ion selective electrode Neodymium PVC-membrane |