Abstract:
A new multisyringe flow injection system for inorganic arsenic determination at trace levels by hydride generation-atomic fluorescence spectrometry (HGAFS) is presented. Preconcentration on a solid-phase was carried out using a column packed with an anion-exchange resin (Amberlite IRA-410). The reagents are dispensed to the system using a multisyringe burette coupled with two multi-port selection valves.

Different parameters were changing in order to make the system as effective as possible. An analytical curve was obtained for arsenic determination between 50 and 2000 ng l−1. This new approach improved five times the sensitivity over a MSFIA–HGAFS technique developed previously by the authors. Detection limit of the proposed technique was (3σb/S) of 30 ng l−1. The relative standard deviation (R.S.D.) of As at 1 μg l−1 was 4.8% (n=7). A sample throughput of 10 h−1 has been achieved. The proposed method has been applied to different reference solid and water materials with satisfactory results.

Keywords:Multisyringe flow injection analysis (MSFIA)   Arsenic   Preconcentration   Hydride generation   Atomic fluorescence