首页 | 本学科首页   官方微博 | 高级检索  
     


Calibration of a compact XUV soft X‐ray monochromator with a digital autocollimator in situ
Authors:Jih-Young Yuh  Shang-Wei Lin  Liang-Jen Huang  Long-Life Lee
Affiliation:National Synchrotron Radiation Research Center, 101 Hsin-Ann Road, Hsinchu Science Park, Hsinchu30076, Taiwan
Abstract:A digital autocollimator of resolution 0.1 µrad (0.02 arcsec) serves as a handy correction tool for calibrating the angular uncertainty during angular and lateral movements of gratings inside a monochromator chamber under ultra‐high vacuum. The photon energy dispersed from the extreme ultraviolet (XUV) to the soft X‐ray region of the synchrotron beamline at the Taiwan Light Source was monitored using molecular ionization spectra at high resolution as energy references that correlate with the fine angular steps during grating rotation. The angular resolution of the scanning mechanism was <0.3 µrad, which results in an energy shift of 80 meV at 867 eV. The angular uncertainties caused by the lateral movement during a grating exchange were decreased from 2.2 µrad to 0.1 µrad after correction. The proposed method provides a simple solution for on‐site beamline diagnostics of highly precise multi‐axis optical manipulating instruments at synchrotron facilities and in‐house laboratories.
Keywords:metrological instrumentation  soft X‐rays  extreme ultraviolet (EUV)  diffraction gratings  autocollimator  on‐site beamline diagnistic
设为首页 | 免责声明 | 关于勤云 | 加入收藏