Implementation of ultrafast X‐ray diffraction at the 1W2B wiggler beamline of Beijing Synchrotron Radiation Facility
Authors:Da-Rui Sun  Guang-Lei Xu  Bing-Bing Zhang  Xue-Yan Du  Hao Wang  Qiu-Ju Li  Yang-Fan Zhou  Zhen-Jie Li  Yan Zhang  Jun He  Jun-Hui Yue  Ge Lei  Ye Tao
Institution:1. Beijing Synchrotron Radiation Facility, Institute of High Energy Physics, Chinese Academy of Sciences, 19B Yuquan Road, Beijing100049, People's Republic of China;2. Accelerator Division, Institute of High Energy Physics, Chinese Academy of Sciences, 19B Yuquan Road, Beijing100049, People's Republic of China
Abstract:The implementation of a laser pump/X‐ray probe scheme for performing picosecond‐resolution X‐ray diffraction at the 1W2B wiggler beamline at Beijing Synchrotron Radiation Facility is reported. With the hybrid fill pattern in top‐up mode, a pixel array X‐ray detector was optimized to gate out the signal from the singlet bunch with interval 85 ns from the bunch train. The singlet pulse intensity is ~2.5 × 106 photons pulse?1 at 10 keV. The laser pulse is synchronized to this singlet bunch at a 1 kHz repetition rate. A polycapillary X‐ray lens was used for secondary focusing to obtain a 72 µm (FWHM) X‐ray spot. Transient photo‐induced strain in BiFeO3 film was observed at a ~150 ps time resolution for demonstration.
Keywords:laser pump/X‐ray probe  ultrafast X‐ray diffraction  photo‐induced strain