A Coulometric Analysis Method and an Ion-exclusion Chromatographic Method for the Determination of Antimony(V) in Large Excess of Antimony(III)
Authors:Akiharu Hioki
Institution:National Metrology Institute of Japan (NMIJ), National Institute of Advanced Industrial Science and Technology (AIST).
Abstract:A coulometric analysis method and an ion-exclusion chromatographic method were developed for the determination of antimony(V) in a large excess of antimony(III). Antimony(V) reacted with potassium iodide in a high concentration hydrochloric acid; the liberated iodine was determined by the standard-addition method using coulometrically generated iodine. Using a Dionex ICE-AS1 ion-exclusion column, antimony(V) was eluted with 40 mmol/L sulfuric acid; on the other hand, antimony(III) was strongly retained on the column. The content, expressed as the amount ratio of antimony(V) to antimony(III), was 0.035% in a 10 g/kg antimony(III) solution prepared from an antimony(III) oxide reagent by the coulometric analysis method and 0.036% in a 1 g/kg antimony(III) solution prepared from the same antimony(III) oxide by the ion-exclusion chromatographic method. The results of both methods were in good agreement with each other. The detection limit of antimony(V) in antimony(III) oxide by the former method was 0.004% of antimony(III), and that by the latter method was 0.002% of antimony(III).