Flow‐Injection Pulsed‐Amperometric Determination of Free Glycerol in Biodiesel at a Gold Electrode |
| |
Authors: | Tatielli G. G. Barbosa Eduardo M. Richter Rodrigo A. A. Muñoz |
| |
Affiliation: | Instituto de Química, Universidade Federal de Uberlandia, Av. Jo?o Naves de?ávila 2121, Bloco 1D, 38400‐902, Uberlandia‐MG, Brazil |
| |
Abstract: | This work reports the highly‐sensitive amperometric determination of free glycerol in biodiesel at a gold electrode adapted in a flow‐injection analysis (FIA) cell. The amperometric method involved the continuous application of three sequential pulses to the working electrode (+250 mV, +700 mV, and ?200 mV, for 100 ms each). This sequence of potential pulses eliminated electrode passivation and dramatically increased the analytical signal. The proposed FIA‐amperometric method presented low relative standard deviation between injections (1.5 %, n=15), high analytical frequency (85 h?1), satisfactory recovery values (93–118 %) for spiked samples, wide linear range (from 1 to 300 µmol L?1), and low detection limit (0.5 µmol L?1). |
| |
Keywords: | Biodiesel Glycerol Flow injection analysis Gold electrode Pulsed amperometry |
|
|