Highly Sensitive Neurotransmitters Analysis at Platinum‐Ultramicroelectrodes Arrays
Authors:Fernando C Moraes  Ivana Cesarino  Dyovani Coelho  Valber A Pedrosa  Sergio A S Machado
Institution:1. Instituto de Química de S?o Carlos, Universidade de S?o Paulo, S?o Carlos, SP, Brasil;2. Institute of Bioscience, Department of Chemistry and Biochemistry, UNESP, Botucatu, SP, Brazil
Abstract:An easy way to determine norepinephrine (NE) in biological fluid using a platinum ultramicroelectrode array (Pt‐UMEAs) is described. Issues related to UME electrode surface treatment and characterizations are also addressed. At optimized experimental conditions the dynamic concentration range was 1.0 to 10.0 µmol L?1 with a detection limit of 40.5 nmol L?1. The repeatability of current responses for injections of 5 µmol L?1 NE was evaluated to be 4.0 % (n=10). This approach obtained excellent sensitivity, a reliable calibration profile and stable electrochemical response for norepinephrine detection. The content of NE in urine samples without any preconcentration, purification, or pretreatment step, was successfully analyzed by the standard addition method using the Pt‐UMEAs.
Keywords:Platinum ultramicroelectrode array  Norepinephrine  Biological fluids