首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of 8[(3S)-3-amino-1-pyrrolidinyl]-1-cyclopropyl-7-fluoro-9-methyl-4-oxo-4H-quinolizine-3-carboxylic acid hydrochloride and related substances by high performance liquid chromatography
Authors:L Elrod Jr  C L Linton  J Morley  T G Golich  C Gay
Institution:(1) Physical Analytical Chemistry Department, Abbott Laboratories, 60064-4000 North Chicago, Illinois, USA
Abstract:Summary The purity of the investigational antibacterial Abbott-086719.1 (I) which is 8(3S)-3-amino-1-pyrrolidinyl]-1-cyclopropyl-7-fluoro-9-methyl-4-oxo-4H-quinolizine-3-carboxylic acid hydrochloride is determined using gradient elution HPLC. The chromatographic separation was optimized and the chromatographic parameters critical to separation are discussed in detail. Most acceptable separations were achieved using an Alltima C18 column (5 mgrm) measuring 4.6 mm I.D.×15 cm with 0.01M–0.03M citrate eluents which were modified with acetonitrile or with mixtures of acetonitrile and methanol. Related substances in I were determinable to 0.05%. Repeatability (RSD values) for determining related substances at levels of 0.50 to 0.06% ranged from ±1.9 to ±10%. Determinations of I in 5% dextrose in water and in hydroxypropyl methylcellulose were made using the same column and a simple isocratic system. The determination of I was stability indicating with precision (RSD values) of ±0.5% to ±2.0% and good agreement with theory for formulations containing I at 1.0 to 95 mg/mL concentrations. Recoveries of I from the vehicles were quantitative and linearity of the detector response of I was demonstrated to at least 0.10 mg/mL.
Keywords:Column liquid chromatography  Fluoroquinolones  Quinolizines  Impurity Profile  Abbott compound 086719  1
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号