首页 | 本学科首页   官方微博 | 高级检索  
     


The influence of background power on the performance of an ideal bolometer
Authors:M. J. Griffin and W. S. Holland
Affiliation:(1) Queen Mary College, University of London, Mile End Road, E1 4NS London, UK
Abstract:
The effects of radiant power loading on the performance of an ideal bolometer are analysed. The operation of the bolometer is characterized in terms of dimensionless parameters which correspond to the temperature sensitivity of the thermometric material, the electrical bias point and the amount of optical power loading. Expressions are derived for the zero frequency NEP and responsivity. Examples are given for the case of a bolometer cooled to 100 mK and used for ground-based astronomical photometry at submillimetre wavelengths.
Keywords:Bolometers  Submillimetre
本文献已被 SpringerLink 等数据库收录!