Evaluation and application of bismuth as an internal standard for the determination of lead in wines by simultaneous electrothermal atomic absorption spectrometry
Authors:Fernandes Kelly G  de Moraes Mercedes  Neto José A Gomes  Nóbrega Joaquim A  Oliveira Pedro V
Affiliation:Departamento de Química Analítica, Universidade Estadual Paulista, Araraquara, SP, Brazil.
Abstract:A method for the direct determination of Pb in wines by simultaneous multi-element atomic absorption spectrometry (SIMAAS) using a transversely heated graphite atomizer, Zeeman-effect background corrector and internal standardization is proposed. Bismuth was used as an internal standard and Pd(NO3)2 plus Mg(NO3)2 as chemical modifier to stabilize both the analyte and the internal standard. The implementation of two pyrolysis steps avoided any build-up of carbonaceous residues on the graphite platform. All diluted samples (1 + 1 v/v) in 0.2% v/v HNO3 and reference solutions (5.0-50 microg l(-1) Pb in 0.2% v/v HNO3) were spiked with 25 microg l(-1) Bi. For a 20 microl aliquot dispensed into the graphite tube, a good correlation (r = 0.9997) was obtained between the ratio of the analyte signal to the internal standard signal and the analyte concentration of the reference solutions. The electrothermal behaviour of Pb and Bi in red, white and rosé wines were compared. In addition, absorbance variations due to changes in experimental conditions, such as atomizer temperature, integration time, injected sample volume, radiation beam intensity, graphite tube surface, dilution and sample composition, were minimized by using Bi as internal standard. Relative standard deviations of measurements based on integrated absorbance varied from 0.1 to 3.4% and from 0.5 to 7.3% (n = 12) with and without internal standard correction, respectively. Good recoveries (91-104%) for Pb spikes were obtained. The characteristic mass was 45 pg Pb and the limit of detection based on integrated absorbance was 0.9 microg l(-1) Pb. Internal standardization increased the lifetime of the tube by 25%. Direct determinations of Pb in wines with and without internal standardization approaches were in agreement at the 95% confidence level. The repeatability and the tube lifetime were improved when using Bi as internal standard. The improvement in accuracy using an internal standard was only observed when the analytical results were affected by errors.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号