End-point sharpness in thermometric titrimetry
Authors:Tyrrell H J
Affiliation:Department of Chemistry, Chelsea College of Science and Technology, Manresa Road, London S.W.3 U.K.
Abstract:
It is shown that the sharpness of an end-point in a thermometric titration where the simple reaction A + B right harpoon over left harpoon AB takes place, depends on Kc(A') where K is the equilibrium constant for the reaction, and c(A') is the total concentration of the titrand (A) in the reaction mixture. The end-point is sharp if, (i) the enthalpy change in the reaction is not negligible, and (ii) Kc(A') > 10(3). This shows that it should, for example, be possible to titrate 0.1 M acid, pK(A) = 10, using a thennometric end-point. Some aspects of thermometric titrimetry when Kc(A') < 10(3) are also considered.