Abstract: | Tunable diode laser absorption spectroscopy (TDLAS) has been widely employed in atmospheric trace gases detection. The ratio of the second-harmonic signal to the intensity of laser beam incident to the multi-pass cell is proved to be proportional to the product of the path length and the gas concentration under any condition. A new calibration method based on this relation in TDLAS system for the measurement of trace gas concentration is proposed for the first time. The detection limit and the sensitivity of the system are below 110 and 31ppbv (parts-per-billion in volume), respectively. |