Band splitting in overloaded isocratic elution chromatography I. The experimental evidence
Authors:Gritti Fabrice  Guiochon Georges
Institution:Department of Chemistry, The University of Tennessee, Knoxville, TN 37996-1600, USA.
Abstract:A series of samples of increasing volume (from 0.001 to 4.0 cm3), containing the same constant concentration (40 g/l) of two simple compounds, ethylbenzoate and 4-tert.-butylphenol were injected on a Kromasil-C18 column with methanol-water (62:38. v/v) as the mobile phase. Complex band profiles were observed when the volume of the sample became large enough and strong band interference took place. The analysis of the fractions collected during the elution of the mixed band demonstrates that, for samples larger than 2 cm3, the band of 4-tert.-butylphenol is split into two separate bands, one eluted before and the other eluted after the band of ethylbenzoate. Such a phenomenon has never been observed yet in RPLC, under isocratic elution conditions.
Keywords:Band splitting  Band profile  Fraction collection  Phenol