Oscillatory disturbance in force calibration of optical tweezers |
| |
Authors: | Liu Chun-Xiang Guo Hong-Lian Jiang Yu-Qiang Li Zhao-Lin Cheng Bing-Ying Zhang Dao-Zhong |
| |
Affiliation: | Optical Physics Laboratory, Institute of Physics and Centre for Condensed Matter Physics, Chinese Academy of Sciences, Beijing,100080, China; State Key Laboratory of Quantum Optics and Quantum Optics Devices, Department of Electronics and Information Technology,Shanxi University, Taiyuan,030006, China |
| |
Abstract: | In the calibration of the optical trap stiffness, it is found that there appears an attenuating oscillation as an oscillatory disturbance added tothe trapped bead movement, when the scanner is driven by a triangular waveinput. An equivalent oscillator model is put forward to explain themechanism of the oscillatory disturbance. Both the measurements andcalculations show that the attenuating oscillation comes from theoscillation of the scanner and the triangular wave drive causes thisadditional oscillation of the scanner. Furthermore, the analysis indicatesthat the oscillatory disturbance will become stronger, when the stiffness ofthe trap increases or the natural frequency of the scanner decreases. Weadopt another driving way, i.e. a sinusoidal wave input is used instead ofthe triangular wave input. Our experiment has verified that in this case theoscillatory disturbance is eliminated completely. |
| |
Keywords: | optical tweezers force calibration oscillatory disturbance |
本文献已被 维普 等数据库收录! |
| 点击此处可从《中国物理》浏览原始摘要信息 |
|
点击此处可从《中国物理》下载免费的PDF全文 |
|