Direct determination of cadmium in high-purity metals by flame atomic fluorescence spectrometry |
| |
Authors: | P. Murugaiyan S. Natarajan Ch. Venkateswarlu |
| |
Affiliation: | Analytical Chemistry Division, Bhabha Atomic Research Centre, Bombay-400 085 India |
| |
Abstract: | Raschig rings used as a safety mechanism, to avoid critical reactions in solutions containing radioactive materials, are usually made of borosilicate glass. Since boron is the active neutron absorbing ingredient, it is important to determine the boron content in the Raschig rings at any given time. A method has been developed to determine rapidly the boron content of borosilicate glasses. Ion exchange and potentiometric measurement are used to determine boron as the tetrafluoroborate ion. The precision of the method is ±2.0 mV. The average difference between values of a wet chemical analysis and those of the potentiometric method is 7.7%. |
| |