首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Sensitivity of a planar micro-flame ionization detector
Authors:Winfred Kuipers  Jörg Müller
Institution:Hamburg University of Technology, Institute of Microsystems Technology, Eissendorfer Str. 42, 21073, Hamburg, Germany
Abstract:The sensitivity of a MEMS μFID with reduced fuel gas consumption for portable applications like mobile GC or THA is examined. It is shown that sensitivity depends on flame size and type of sample gas supply (either separate supply or premixed with the hydroxygen). In contrast to conventional FIDs, the sensitivity of the μFID increases with decreasing molecule size. The sensitivity to methane can be optimized up to conventional values. Measurements with the μFID as a second detector in a μGC module prove the additional functionality of such a system.
Keywords:Micro-flame ionization detector  MEMS  Microsystems technology  Micro-gas chromatography  Total hydrocarbon analysis
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏