首页 | 本学科首页   官方微博 | 高级检索  
     


Accuracy limits for planar measurements of flow field velocity, temperature and pressure using Filtered Rayleigh Scattering
Authors:J. N. Forkey  W. R. Lempert  R. B. Miles
Affiliation:(1) Princeton University Mechanical & Aerospace Engineering Department Room D-414 Engineering Quadrangle Olden St. Princeton, NJ 08544, USA, US
Abstract:
 We present experimental results using Filtered Rayleigh Scattering to make planar measurements of velocity, temperature and pressure in ambient air and in a Mach 2 free jet. The ambient air measurements are used to identify and calibrate experimental uncertainties. The Mach 2 measurements demonstrate the usefulness of the technique for making accurate planar measurements in a high speed flow. The measured values for velocity, temperature and pressure in the Mach 2 jet ranged, through a shock system, from 205 to 235 m/s, 150 to 170 K and 700 to 1000 torr, with estimated uncertainties of ±5.4 m/s, ±3.2 K and ±38 torr (±2 to 3%, ±2% and ±4–5%, respectively). Received: 10 December 1996/Accepted: 14 July 1997