Voltammetric determination of Se(IV) and Se(VI) in saline samples—Studies with seawater, hydrothermal and hemodialysis fluids
Authors:Paulo C do Nascimento  Cristiane L Jost  Denise Bohrer
Institution:a Departamento de Química, Universidade Federal de Santa Maria, C.P. 5051, 97105-970 Santa Maria, RS, Brazil
b School of Engineering and Science, Geosciences and Astrophysics, Jacobs University Bremen GmbH, P.O. Box 750561, D-28725 Bremen, Germany
Abstract:Determination of Se(IV) and Se(VI) in high saline media was investigated by cathodic stripping voltammetry (CSV). The voltammetric method was applied to assay selenium in seawater, hydrothermal and hemodialysis fluids. The influence of ionic strength on selenium determination is discussed. The CSV method was based on the co-electrodeposition of Se(IV) with Cu(II) ions and Se(VI) determined by difference after sample UV-irradiation for photolytic selenium reduction. UV-irradiation was also used as sample pre-treatment for organic matter decomposition. Detection limit of 0.030 μg L−1 (240 s deposition time) and relative standard deviation (RSD) of 6.19% (n = 5) for 5.0 μg L−1 of Se(IV) were calculated. Linear calibration range for selenium was observed from 1.0 to 100.0 μg L−1. Concerning the pre-treatment step, best results were obtained by using 60 min UV-irradiation interval in H2O2/HCl medium. Se(VI) was reduced to the Se(IV) electroactive species with recoveries between 91.7% and 112.9%. Interferents were also investigated.
Keywords:Selenium  Ionic strength  Speciation analysis  Cathodic stripping voltammetry  UV-irradiation