首页 | 本学科首页   官方微博 | 高级检索  
     检索      


On-line determination of trace levels of palladium by flame atomic absorption spectrometry
Authors:Iglesias Mònica  Anticó Enriqueta  Salvadó Victòria
Institution:Dept. Quimica, Universitat de Girona, Campus Montilivi, Girona, Spain.
Abstract:A preconcentration method is developed for the on-line determination of palladium in complex matrices with flame atomic absorption spectrometry (FAAS). The flow system comprised of a minicolumn filled with polyamine Metalfix-Chelamine resin which is highly selective for Pt(IV), Au(III) and Pd(II). Best preconcentration conditions are established by testing different resin quantities, sample and eluent solution volumes, and adsorption and elution steps flow rates. Sample volumes of 4.7 ml of palladium solutions resulted in an enrichment factor of twenty at the optimum hydrodynamic conditions. This value can be increased by injecting larger volumes of sample solution. The method is sensitive, easy to operate and permitted the determination of sub-mg l−1 levels of palladium with a detection limit of 0.009 mg l−1. The resin was used up to 60 times in consecutive retention-elution cycles without any appreciable deterioration in its performance. The applicability of this method was tested by determining the palladium content in synthetic geological samples as well as in the pellet-type used car catalyst reference material.
Keywords:Pd(II)preconcentration  On-line sub-mg l&minus  1 determination  Flame atomic absorption spectrometry  Metalfix-Chelamine resin
本文献已被 ScienceDirect PubMed 等数据库收录!