首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of gold in ore by flame atomic absorption spectrometry with flow-injection on-line sorbent extraction preconcentration
Authors:Xu Shukun  Sun Lijing  Fang Zhaolun
Affiliation:Flow Injection Analysis Research Centre, Institute of Applied Ecology, Academia Sinica, Shenyang China
Abstract:Gold in ores was determined by flame atomic absorption spectrometry following on-line preconcentration by sorbent extraction in a flow-injection system. The medium polarity adsorption resin Amberlite XAD-8 packed in a 220-μl micro-column was used to collect gold(III) from hydrochloric acid sample solutions for 40 s at 7.6 ml/min. Ethanol was used to elute the adsorbed analytes into the nebulizer. Optimization studies were made on sample loading rate, elution rate and sample acidity. Some possible interferences on the determination are discussed. A 35-fold enrichment was achieved at a sampling frequency of 60 h?1 and with an RSD of 1.4%. The detection limit (3σ) and 2 μg l?1. Results for gold in ore samples showed good agreement with those obtained using activated carbon adsorption preconcentration. The recoveries were 97–108%.
Keywords:Atomic absorption spectrometry  Flow system  Gold  Ores  Preconcentration
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号