Microdetermination of fluorine in organic compounds by oxygen flask combustion and conductometric titration |
| |
Authors: | Akio Nara Nobuko Kobayashi Kazuko Hasegawa Katsuko Honba Shigeo Baba |
| |
Institution: | 1. Tokyo Research Laboratories, Kowa Company, Ltd. Noguchi-cho, Higashimurayama-city, Tokyo, Japan;2. Tokyo College of Pharmacy, Kitashinjuku, Tokyo, Japan |
| |
Abstract: | For the conductometric titration of fluoride with calcium acetate, solvent system in sample solution and titrant, concentration, and acidity of sample solution were examined to establish the titration conditions. Results of these examinations were transferred to the microdetermination of fluorine in organic compounds using oxygen flask combustion method.Comparative examination between quartz and Pyrex flasks for the combustion of fluorine samples indicated that good results were obtained by the use of the former flask, whereas the use of the latter one gave negative values of 1.0–1.6% due to the formation of boron fluoride during the combustion. |
| |