Simultaneous determination of tin and lead by a.c. anodic stripping voltammetry at a hanging mercury drop electrode sensitized by cetyltrimethylammoniu |
| |
Authors: | JHernández Méndez RCarabias Martínez MEGonzález López |
| |
Institution: | Department of Analytical Chemistry, University of Salamanca Spain |
| |
Abstract: | The influence of cetyltrimethylammonium bromide (CTAB) on the simultaneous determination of tin(IV) and lead(II) by anodic stripping voltammetry at a hanging mercury drop electrode (HMDE) in a 0.1 M hydrochloric acid—0.1 M oxalic acid medium was studied using d.c. and a.c. stripping. In the presence of CTAB, tin and lead show voltammetric peaks separated by 100 mV, the sensitivity depending on the concentration of CTAB. The best conditions for the simultaneous determination of both elements (2 × 10-3 M) were found. A method is proposed for the determination of tin in the presence of lead and three procedures are given for the determination of lead in the presence of tin. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |