Determination of anabolic steroids by gas chromatography/negative-ion chemical ionization mass spectrometry and gas chromatography/negative-ion chemical ionization tandem mass spectrometry with heptafluorobutyric anhydride derivatization
Authors:Choi M H  Chung B C  Lee W  Lee U C  Kim Y
Affiliation:Doping Control Centre, Korea Institute of Science and Technology, Cheongryang, Seoul, Korea.
Abstract:A gas chromatography/mass spectrometry (GC/MS) method is described which uses negative ion chemical ionization (NCI) and tandem mass spectrometry (MS/MS) for the determination of eight anabolic steroids in human urine. Eight anabolic steroids were derivatized by heptafluorobutyric anhydride (HFBA), and were determined using GC/NCI-MS and GC/NCI-MS/MS. The linear correlation coefficients for calibration in NCI-MS/MS were in the range 0.9880-0.9988. This method of derivatization with HFBA for use with GC/NCI was useful in determinations of 19-norandrosterone, boldenone, 19-noretiocholanolone, 2-methylandrosterone, nandrolone, 1-methyleneandrosterone, 1-methylandrosterone, 4-dihydroboldenone and mesterolone. The detection limits of this procedure were 5-20 ppb at a signal-to-noise (S/N) ratio of 3.