Identification of137Cs in an individual microparticle by laser desorption/ionization ion trap mass spectrometer |
| |
Authors: | KwangHee Hong KyuSeok Song HyungKi Cha Mo Yang JongMin Lee ChangWo Lee GaeHo Lee |
| |
Affiliation: | (1) Laboratory for Quantum Optics, Korea Atomic Energy Research Institute, 305-600 Taejon, Korea;(2) Environmental Radiation Assessment Division, Korea Atomic Energy Research Institute, 305-600 Taejon, Korea;(3) Department of Chemistry, Chungnam National University, 305-764 Taejon, Korea |
| |
Abstract: | The detection of radiocesium in microparticles was performed by using an ion trap mass spectrometer coupled with laser desorption and ionization. Pulsed laser desorbed particle and the resulted ions were analyzed by an ion trap mass analyzer. The presence of radiocesium, especially about137Cs, in microparticles was verified by single as well as successive particle analysis. The detection limit was reached to ≈ag/particle level with a signal-to-background ratio of 4. The inhomogeneous distribution of particle size and the irregular shapes of particle limit the quantitative evaluation of137Cs concentration in the microparticle. But this high sensitivity allows to monitor directly the radiocesium from small amounts of a microparticle sample. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|