A flow-based enzyme-linked immunosorbent assay on a polydimethylsiloxane microchip for the rapid determination of immunoglobulin A |
| |
Authors: | Nakajima Hizuru Yagi Maiko Kudo Yuki Nakagama Tatsuro Shimosaka Takuya Uchiyama Katsumi |
| |
Affiliation: | a Faculty of Urban Environmental Sciences, Tokyo Metropolitan University, 1-1 Minamiohsawa, Hachioji, Tokyo 192-0397, Japan b National Metrology Institute of Japan (NMIJ), National Institute of Advanced Industrial Science and Technology (AIST), Tsukuba Central 3, Umezono, Tsukuba, Ibaraki 305-8563, Japan |
| |
Abstract: | A flow-based enzyme-linked immunosorbent assay (ELISA) on a polydimethylsiloxane (PDMS) microchip has been developed for the rapid determination of immunoglobulin A (IgA). The analytical principle of this integrated method is the same as the conventional sandwich-type ELISA. A primary antibody (anti-IgA) was adsorbed on the surface of a PDMS microchannel, and then an antigen (IgA) and a secondary antibody (anti-IgA HRP labeled) were reacted successively. The resulting antigen-antibody complex, fixed on the surface of the microchannel, was detected using Amplex® Red and a fluorescent imaging system. The calibration curve of the IgA standard solution was linear in the range of 0-50 ng/mL at the flow rate of 10 μL/min. This flow rate corresponds to the reaction time of 4.8 s. Compared to the conventional assay on a 96-well microtiter plate, the present assay on the microchip dramatically shortened the reaction time necessary for the enzyme-substrate reaction from 30 min to 4.8 s, i.e., to 1/375. The amounts of the reagent and sample were also reduced to 1/100 compared to the 96-well microtiter plate. |
| |
Keywords: | Microchip ELISA Rapid determination Immunoglobulin A Saliva Stress |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|