首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Flameless atomic-absorption determination of phosphorus using ZrC coated graphite atomizer tubes
Authors:I Havezov  E Russeva and N Jordanov
Institution:(1) Institute of General and Inorganic Chemistry, Bulgarian Academy of Sciences, 1040 Sofia, Bulgaria
Abstract:Summary A flameless atomic-absorption method for determination of phosphorus is developed, employing zirconium treated graphite tubes. The great improvement of the sensitivity with these tubes eliminates the necessity of adding La(NO3)3. The sensitivity does not vary from tube to tube and its maximal value is attained within the first 2–3 firings. The atomization signal is practically the same within the range of 2,600–2,800° C when the rise of the temperature is high (> 1,500° C/s). The best signal to noise ratio is obtained at 2,600° C. The detection limit is 0.05 ppm with injection of 50 mgrl. The method is applied to the analysis of a standard steel sample.
Flammenlose Atomabsorptionsbestimmung von Phosphor mit einer ZrC-beschichteten Graphitrohrküvette
Zusammenfassung Durch Verwendung einer ZrC-beschichteten Graphitrohrküvette wurde eine wesentliche Erhöhung der Empfindlichkeit erzielt, ohne daß eine Zugabe von La(NO3)3 nötig wäre. Bei Küvettenwechsel ändert sich die Empfindlichkeit nicht. Das maximale Signal wird während der ersten 2–3 Bestimmungen erreicht. Das Atomisierungssignal bleibt im Bereich zwischen 2600 und 2800 °C praktisch unverändert. Das Signal/Rausch-Verhältnis ist bei 2600 °C am größten. Die Nachweisgrenze beträgt 0,05 ppm bei 50 mgrl. Die Methode wurde zur Analyse von Stahlstandardproben angewendet.
Keywords:Best  von Phosphor in Stahl  Spektralphotometrie  Atomabsorption  flammenlos  ZrC-beschichtete Graphitrohrkü  vette
本文献已被 SpringerLink 等数据库收录!