Correlation of urokinase activity from biopotency and high-performance liquid chromatographic assays
Authors:R A Cox  K N McFarland  P H Sackett  M T Short
Abstract:
A simpler, less expensive, and faster high-performance liquid chromatographic method was shown to be an alternative to urokinase potency determinations by the Ploug method. Post-elution recovery of the low-molecular-weight form was 104 +/- 2.4% as determined by the Ploug method. Two analysts reported relative standard deviations of 1.6% and 1.1% based on peak height determination of eight replicate injections of a single sample of low-molecular-weight material. Linearity at the same wavelength for low- and high-molecular-weight forms was 0.9999 and 0.9992, respectively, for peak height versus potency.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏