首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A catalytic immuno-reactor for the amperometric determination of human serum albumin
Authors:Isao Karube  Tadashi Matsunaga  Takeru Satoh  Shuichi Suzuki
Institution:Research Laboratory of Resources Utilization, Tokyo Institute of Technology, Nagatsuta-cho, Midori-ku, Yokohama 227 Japan
Abstract:A catalytic immuno-reactor for the determination of human serum albumin (HSA) was constructed by using immobilized antibody and an amperometric detector. A sandwich assay with hemin-labeled antibody to catalyze the decomposition of hydrogen peroxide was used, the catalytic activity of the hemin-antibody conjugate being determined by measuring the decrease in hydrogen peroxide concentration. The reaction of hemin-labeled antibody with antigen was complete within 30 min and the current decrease was correlated with the HSA concentration. The relative standard deviation was about 9% at an HSA concentration of 1 mg ml?1.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏