Simultaneous determination of hypoxanthine,inosine, inosine-5′-phosphate and adenosine-5′-phosphate with a multielectrode enzyme sensor |
| |
Authors: | Etsuo Watanabe Shunsuke Tokimatsu Kenzo Toyama Isao Karube Hideaki Matsuoka Shuichi Suzuki |
| |
Affiliation: | Department of Food Engineering and Technology, Tokyo University of Fisheries, 4-5-7 Konan, Minato-ku, Tokyo 108 Japan;Research Laboratory of Resource Utilization, Tokyo Institute of Technology, Nagatsuta-cho, Midori-ku, Yokohama 227 Japan |
| |
Abstract: | A multielectrode enzyme sensor for the simultaneous determination of adenosine-5′-phosphate (AMP), inosine-5′-phosphate (IMP), inosine (HXR) and hypoxanthine (HX)in fish meat was developed by assembling four enzyme sensors for AMP, IMP, HXR and HX in a flow cell. These compounds were determined from oxygen consumption according to the following reactions: AMP IMP HXR HX Uric acid where AD is AMP deaminase, NT is 5′-nucleotidase, NP is nucleoside phosphorylase and XO is xanthine oxidase. Enzymes were covalently bound to a membrane prepared from cellulose triacetate, 1,8-diamino-4-aminomethyloctane and glutaraldehyde. Sensors for HX, HXR, IMP and AMP were prepared by attaching membranes of XO, XONP, XO NPNT, and of XONPNT and AD, respectively, to four oxygen electrodes. Samples extracted from sea bass, bream, flounder, abalone and arkshell were analyzed within 5 min, from the simultaneous response curves of the four electrodes. Results obtained by the multisensor system were in good agreement with those determined by each single electrode. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |
|