首页 | 本学科首页   官方微博 | 高级检索  
     


The determination of trace amounts of heavy metals in waters by a flow-injection system including ion-exchange preconcentration and flame atomic absorption spectrometric detection
Authors:Fang Zhaolun  Xu Shukun  Zhang Suchun
Affiliation:Institute of Forestry and Soil Science, Academia Sinica, Shenyang China
Abstract:
The flow-injection system combines on-line ion-exchange preconcentration with atomic absorption spectrometry (a.a.s.) for the determination of traces (μg l?1) of heavy metals in water samples. Miniature columns packed with 60–100 mesh chelating resin (122) with a salicylic acid functional group are used for preconcentration. A multifunctional rotary sampling valve which incorporated two parallel sampling columns allows sampling, exchange, elution and a.a.s. to be achieved sequentially. The increases in sensitivity for nickel, copper, lead and cadmium were 20–28-fold at a sampling rate of 40 h?1 with 5-ml samples. Relative standard deviations were 1.5–4.1%. The recoveries of these four metals added to tap, sea and polluted waters were generally satisfactory, except for cadmium in polluted water. The effects of column diameter and elution flow rates on sensitivity are discussed. Possible interferences are described.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏