A selective electrolytic sensor for nitrite based on a modified platinum electrode
Authors:James A Cox  Pawel J Kulesza
Institution:Department of Chemistry and Biochemistry, Southern Illinois University, Carbondale, IL 62901 U.S.A.
Abstract:The electrolytic sensor described is based on the oxidation of nitrite at a platinum electrode modified with chemisorbed iodine and coated with a thin layer of quaternized poly(4-vinylpyridine), qPVP. The sealed sensor uses an anion-exchange membrane to separate Donnan transport of nitrite across the membrane and controlled potential electrolysis at the Pt/qPVP indicator electrode. The sensor has a linear response to nitrate concentration in aqueous samples over the range 4 × 10?6?2 × 10?3 M nitrite. The detection limit is 2 × 10?6 M nitrite. The sensor is free of interference by nitrate, dissolved oxygen, cations, and many neutral species. Anions that are electroactive at 0.7 V vs. Ag/ AgCl would interfere, but they are uncommon in most samples. Initial tests with lake water samples suggest that this sensor is unaffected by this matrix. The system was also evaluated for monitoring nitrite levels in spiked meat extracts.