Chemiluminescence determination of carbofuran at trace levels in lettuce and waters by flow-injection analysis |
| |
Authors: | Huertas-Pérez José Fernando Gámiz-Gracia Laura García-Campaña Ana M González-Casado Antonio Martínez Vidal José L |
| |
Affiliation: | a Department of Analytical Chemistry, Faculty of Sciences, University of Granada, E-18071 Granada, Spain b Department of Hydrogeology and Analytical Chemistry, Faculty of Experimental Sciences, University of Almería, E-04120 Almería, Spain |
| |
Abstract: | A simple, fast chemiluminescence (CL) flow-injection (FI) method based on the reaction of luminol with KMnO4 in alkaline medium has been described for the direct determination of carbofuran. The method is based on the enhancing effect in the emission light from the oxidation of luminol produced in presence of carbofuran. The optimisation of instrumental and chemical variables influencing the CL response of the method has been carried out by applying experimental design, using the proposed flow-injection manifold. Under the optimal conditions, the CL intensity was linear for a carbofuran concentration over the range of 0.06-0.5 μg ml−1, with a detection limit of 0.02 μg ml−1. The method has been successfully applied to the determination of carbofuran residues in spiked water and lettuce samples. |
| |
Keywords: | Carbofuran Chemiluminescence Luminol Flow-injection analysis |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|