首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A realization of digital wireless transmission for MRI signals based on 802.11b
Authors:Wei Juan  Liu Zhengguang  Chai Zhi  Yuan Jing  Lian Jianyu  Shen Gary X
Institution:Department of Electrical and Electronic Engineering, The University of Hong Kong, Pokfulam, Hong Kong.
Abstract:In this paper, a digital wireless transmission system based on 802.11b standard for magnetic resonance imaging (MRI) application is designed and built for the first time to eliminate the interference aroused by coil array cables. The analysis shows that the wireless receiver has a very high sensitivity to detect MRI signals. The modulation technique of differential quadrature phase shift keyed (DQPSK) can be applied to MRI data transmission with rate of 2 Mbps and bandwidth of 2 MHz. The bench test verifies that this wireless link has a dynamic range over 86 dB supporting up to 3 T MRI system data transmission. The 2D spin echo imaging of phantom is performed and the SNR of the image obtained by the wireless transmission can be comparable with that got by the coaxial cables.
Keywords:Magnetic resonance imaging (MRI)  Digital wireless transmission  Coil array  IEEE 802  11
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏