Determining a 'safe' high-mass limit in matrix-assisted laser desorption/ionisation time-of-flight mass spectra of coal derived materials with reference to instrument noise
Authors:MJ Lazaro  AA Herod  M Domin  Y Zhuo  CA Islas  R Kandiyoti
Institution:Department of Chemical Engineering and Chemical Technology, Imperial College (University of London), London SW7 2BY, UK.
Abstract:Three methods for determining a 'safe' estimate for high-mass limits of MALDI spectra of coal derived liquids were explored, using a sample of coal-tar pitch and its pyridine-insoluble fraction. Co-addition of increasing numbers of single-shot spectra (10, 30, 50 and 100 pulses) showed visually observable reductions in noise levels, consistent with robust and statistically meaningful signals. Three separate types of post-acquisition calculation were used to identify high-mass limits of the spectra. (i) A literature method indicated high-mass limits similar to those observed visually-as a shift from baseline at the highest masses, nearly 350 000 u for the coal tar pitch and about 390 000 u for its pyridine insoluble fraction. (ii) Comparing instrument signal with pre-selected multiples of the standard deviation, upper mass estimates of between 40-60 000 u for the coal-tar pitch and about 95 000 u for its pyridine-insoluble fraction were found. (iii) Calculation of the slope was used to identify 'lift-off' of the spectrum from baseline. The angle between the smoothed spectrum and the baseline was matched to a pre-selected value (e.g. 0.5 degrees and 1 degrees ). However, the arbitrary specification of the key parameter did not establish this last method on a firm basis. The choice of a criterion for estimating high-mass limits of MALDI spectra remains a semi-quantitative procedure; a reasonably conservative high-mass limit may be estimated by comparison of signal with five times the standard deviation. However, evaluation of size exclusion chromatograms of the present samples using polystyrene standards suggests that molecular mass distributions of pitch samples arrived at by MALDI mass spectrometry are, at least partly, determined by the limitations of available instruments. Copyright 1999 John Wiley & Sons, Ltd.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号