Screening for anabolic steroids in doping analysis by liquid chromatography/electrospray ion trap mass spectrometry |
| |
Authors: | Deventer K Eenoo P Van Delbeke F T |
| |
Affiliation: | Ghent University (UGent), Department of Clinical Chemistry, Microbiology and Immunology, Doping Control Laboratory (DoCoLab), Technologiepark 30, B-9052 Zwijnaarde, Belgium. Koen.Deventer@UGent.be |
| |
Abstract: | A fast and selective LC/MS/MS method for the screening of four anabolic steroids in human urine has been developed and validated. Liquid-liquid extraction with diethyl ether was applied after enzymatic hydrolysis. Analyses were performed on an ion trap mass spectrometer equipped with electrospray ionisation. MS/MS was applied for all compounds. The analytical run time was 11 min. The LOD for all compounds varied between 1 and 10 ng/mL. Left-over A samples, which were declared positive by GC/MS for the presence of 3'-hydroxystanozolol, were assessed using the described method. |
| |
Keywords: | anabolic steroids stanozolol 3′‐hydroxystanozolol gestrinone 17α‐trenbolone doping analysis |