首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A novel nitrate-selective electrode based on precipitated nitron nitrate
Authors:Uma Shankar Lal  M C Chattopadhyaya  Arun K Dey
Institution:(1) Chemical Laboratories, University of Allahabad, India;(2) Department of Chemistry, University of Allahabad, 211002 Allahabad, India
Abstract:Summary A nitrate-selective electrode based on precipitated nitron nitrate in an epoxy resin matrix has been prepared. The lower detection limit is 10–5 M, the optimum pH range is 2.0–8.5, and the response time is 1 min. The electrode is selective in presence of CH3COO, SO4 2–, C2O4 2–, PO4 3–, NO2 , Cl and Br, but I interferes. Results for nitrate in soil samples were in agreement with values obtained spectrophotometrically.
Eine neue nitrat-spezifische Elektrode auf der Basis gefällten Nitronnitrats
Zusammenfassung Eine nitrat-spezifische Elektrode auf der Basis gefällten Nitronnitrats in einer Epoxyharz-Matrix wurde hergestellt. Die untere Nachweisgrenze ist 10–5 M, das pH-Optimum liegt zwischen 2,0 und 8,5, die Reaktionszeit beträgt 1 Minute. Die Elektrode ist selektiv in Gegenwart von Acetat, Sulfat, Oxalat, Phosphat, Nitrit, Chlorid und Bromid; aber Jodid stört. Ergebnisse der Nitratbestimmung in Bodenproben stimmen mit spektrophotometrischen Resultaten überein.


Presented at the 8th International Microchemical Symposium, Graz, August 25–30, 1980.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号