首页 | 本学科首页   官方微博 | 高级检索  
     


Thermal gradients in differential scanning calorimetry
Authors:M. J. Richardson  P. Burrington
Affiliation:1. Division of Materials Applications, National Physical Laboratory, Teddington, Middlesex, UK
Abstract:A combined static and dynamic temperature calibration is described. The static calibration corrects the instrumental dial temperature reading. The dynamic calibration has instrumental and material components and therefore varies from specimen to specimen. It is obtained from individual DSC curves and so removes uncertainties in sample temperature due to varying mass, geometry, and heating rate. The instrumental performance is improved and specific heats may be obtained to an accuracy of ±1%.
Keywords:
本文献已被 SpringerLink 等数据库收录!