首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of chloroquine in biological material by gas-liquid chromatography (author's transl)]
Authors:A Viala  J P Cano  A Durand
Abstract:A method is developed for the gas chromatographic determination of chloroquine after extraction from biological material, using medazepam as the internal standard and a column filled with 3% OV-17 on Gas-Chrom Q. The detection limits are: 0.15 mug/ml of urine, 0.25 mug/ml of whole blood, and 1.50 mug/g of tissue. The technique can be applied in toxicological analysis and in monitoring drug levels.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏