首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A new hydride generation system applied in determination of arsenic species with ion chromatography-hydride generation-atomic fluorescence spectrometry (IC-HG-AFS)
Authors:Changjin Wei
Institution:Beijing Titan Instruments Co. Ltd., Beijing 100016, China
Abstract:A novel procedure was developed for the determination of arsenite (As(III)), arsenate (As(V)), monomethylarsonic (MMA) and dimethylarsinic acid (DMA) with ion chromatography-hydride generation-atomic fluorescence spectrometry (IC-HG-AFS) by employing a new gas-liquid separator (GLS). The effective separation of the four arsenic species was achieved in about 12 min. With a sample loading volume of 20 μl, the measurable minimum for As(III), DMA, MMA and As(V) were 0.02, 0.045, 0.043 and 0.166 ng, respectively, along with relative standard deviations of 1.1, 1.1, 1.7 and 2.2% at the 100 μg l−1 level (n = 6) for As(III), DMA, MMA and As(V), respectively. The present procedure was applied for the speciation of arsenic in underground water and in urine samples, and the sum of the four arsenic species by IC-HG-AFS was in good agreement with the total value by HG-AFS.
Keywords:Arsenic speciation  Ion chromatography  Hydride generation-atomic fluorescence spectrometry  Gas-liquid separator
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏